How to read this converter
Choose a unit and enter a size, or choose a needle tubing gauge. Simulate loads a labelled conversion example from the reviewed catalog.
Values are rounded for display. The needle-tubing lookup uses the unrounded outside diameter.
French, millimetres and inches
The French size is the outside diameter in millimetres multiplied by 3: one French is exactly one third of a millimetre.
Divide millimetres by 25.4 to obtain inches, or divide the French size by 76.2.
For a circular cross-section, circumference is π times diameter; its numerical value in millimetres is about 4.7% greater than the French number.
What the size means
A nominal French size does not establish a measured device diameter or its inner diameter.
Some tapered catheters are labelled with separate tip and hub French sizes.
In a study of 200 interventional radiology device labels, Kibriya and colleagues reported wide variation in the metric equivalent of French sizing.
Needle tubing gauge
ISO 9626:2016 assigns gauges from 10G to 34G to outside-diameter ranges for rigid stainless-steel needle tubing.
A larger gauge number denotes a smaller outside diameter.
This converter reports both gauges where ranges overlap.
There are 23 internal gaps and one overlap: 29G and 28G overlap from 0.349 to 0.351 mm, inclusive.
The designated metric size is a nominal label, not a substitute for the limits: 31G is designated 0.25 mm but starts at 0.254 mm, and 33G is designated 0.20 mm but starts at 0.203 mm.
IV cannula gauge differs
In ISO 10555-5:2013, the optional 20G designation for an over-needle peripheral IV catheter corresponds to an outside-diameter range of 0.950–1.149 mm; ISO 9626:2016 gives 0.860–0.920 mm for 20G needle tubing.
This converter uses the needle-tubing table only.
Decimal inches
Guidewire sizes are often expressed in thousandths of an inch: 0.018 in means 18 thousandths of an inch and converts to exactly 0.4572 mm.
Worked example
Published conversion example — Ojha 2023, Measurement, PMID 36960563.
3 Fr equals 1.00 mm, approximately 0.039 in.
The converter’s ISO 9626:2016 lookup places this diameter between 20G (0.860–0.920 mm) and 19G (1.030–1.100 mm).
Frequently asked questions
How many millimetres is 1 French?
One French is exactly one third of a millimetre: divide the French size by 3, so 3 Fr is 1 mm.
What is 0.035 inch in mm?
0.035 in is exactly 0.889 mm, equivalent to 2.667 Fr (approximately 2.67 Fr).
How do you convert French to needle gauge?
Gauge is a size designation, not a unit. Divide the catheter’s French size by 3 to get its nominal outside diameter in millimetres, then compare that diameter with the ISO 9626:2016 needle-tubing ranges. 5 Fr is 5/3 mm, approximately 1.67 mm, within the 16G outside-diameter range (1.600–1.690 mm). Some diameters fall between ranges.
Which needle-tubing range contains a 7 Fr outside diameter?
7 Fr is 7/3 mm, approximately 2.33 mm. That diameter lies within the ISO 9626:2016 range for 13G needle tubing: 2.300–2.500 mm.
Is 16 Fr or 14 Fr bigger?
16 Fr represents the larger outside diameter: approximately 5.33 mm, compared with 4.67 mm for 14 Fr.
Why are catheters measured in French?
Charrière was a French maker of medical instruments. UCUM notes that the Charrière unit is called ‘French’ in the United States.
References
- Schadow G, McDonald CJ. The Unified Code for Units of Measure. Version 2.2, 17 June 2024. Section 44. https://ucum.org/ucum
- ISO 9626:2016. Stainless steel needle tubing for the manufacture of medical devices — Requirements and test methods. Clause 5.6, Table 1. Numeric requirements reviewed through the identical GOST R ISO 9626-2020 adoption. https://www.iso.org/standard/60480.html
- Thompson A, Taylor BN. Guide for the Use of the International System of Units (SI). NIST Special Publication 811, 2008 edition. Appendix B. doi:10.6028/NIST.SP.811e2008
- ISO 10555-1:2013 + Amd 1:2017. Informative Annex H. Reviewed through GOST ISO 10555-1-2021; this edition is superseded by ISO 10555-1:2023.
- Ojha V, Raju SN, Deshpande A, Ganga KP, Kumar S. Catheters in vascular interventional radiology: an illustrated review. Diagn Interv Radiol. 2023;29(1):138–145. doi:10.5152/dir.2022.21233. PMID 36960563.
- Kibriya N, Hall R, Powell S, How T, McWilliams RG. French sizing of medical devices is not fit for purpose. Cardiovasc Intervent Radiol. 2013;36(4):1073–1078. doi:10.1007/s00270-013-0557-7. PMID 23354967. Abstract reviewed.
- ISO 10555-5:2013. Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral catheters. Table 1. Reviewed through the identical GOST ISO 10555-5-2021 adoption.